A back table illustration of the technique published by PCRonline 🫀 to remove a stuck perclose footplate in the vessel! Simple and I believe can get us out of trouble! #CardioX peeps, comments welcome

73,535 次观看 • 4 个月前 •via X (Twitter)

22 条评论

Michael Megaly 的头像
Michael Megaly4 个月前

Tried it on the purple one, it works as well @Umair2017

Elliot B. Sambol, MD 的头像
Elliot B. Sambol, MD4 个月前

@PCRonline I’ve had this happen. Often times the foot plate doesn’t retract completely because there’s tissue stuck in the groove where it closes. Sometimes just opening and closing a few times is enough to release it and allow for retraction.

Michael Megaly 的头像
Michael Megaly4 个月前

@PCRonline Yup thats the first thing we should do

Michael Sgroi, MD 的头像
Michael Sgroi, MD4 个月前

@PCRonline You can pop open the back of it and there is a clear plastic piece on the inside. When you pull on this clear piece it will retract the footplate. Sometimes the knot can also be stuck so cut the suture too.

Salman Arain 的头像
Salman Arain4 个月前

Thank you Michael. This is helpful. In our case, the footplate wouldn’t retract, even after we dismantled* the device. We found tissue trapped between the footplate and its housing. FWIW, I took the Perclose apart at the handle - which is how I learned it. Not sure if that made a difference.

Michael Megaly 的头像
Michael Megaly4 个月前

@PCRonline your case is different, very tough situation

Wassim Mosleh, MD 的头像
Wassim Mosleh, MD4 个月前

@PCRonline Wish I knew this earlier!

Michael Megaly 的头像
Michael Megaly4 个月前

@PCRonline We learn every day my friend

Joseph L. Mills MD 的头像
Joseph L. Mills MD4 个月前

@Asiritrauma1 @PCRonline Good Post

Irfan Nazir Hassan 的头像
Irfan Nazir Hassan4 个月前

@ChengaziMD @PCRonline Had it happen to me once where I actually accessed through a CFA stent inadvertently and realized end of procedure. Eventually was able to retrieve it but was scary for sure 😱

Epic Immaculate 的头像
Epic Immaculate4 个月前

@willsuh76 @PCRonline Great Tip! Abbott reps can’t tell you to do this. Why? Guidant reps once instructed doctors to manipulate the Ancure device the same way — it led to the device’s recall. No device company or rep from that era ever forgot the lesson.

Bilal Iqbal 的头像
Bilal Iqbal4 个月前

@PCRonline Great demonstration

Michael Megaly 的头像
Michael Megaly4 个月前

@PCRonline Thank u brother!

Ronnie P 的头像
Ronnie P4 个月前

@PCRonline Great info Dr. Megaly!

Michael Megaly 的头像
Michael Megaly4 个月前

@PCRonline Thank u sir

Sabeeda Kadavath 的头像
Sabeeda Kadavath4 个月前

@ABeneduceMD @PCRonline Good demonstration Prof !

Michael Megaly 的头像
Michael Megaly4 个月前

@ABeneduceMD @PCRonline Lol!! glad you liked it

Charles Beale MD 的头像
Charles Beale MD4 个月前

@PCRonline I find it easier to insert a kelly into the top of perclose(light blue part) open it, take off black lever and release the footplate

Guillermo Escobar🩸 的头像
Guillermo Escobar🩸4 个月前

@RMotaganahalli @PCRonline Do you know if this works on the purple one?

Michael Megaly 的头像
Michael Megaly4 个月前

@RMotaganahalli @PCRonline Will try it at the back table

calif 的头像
calif4 个月前

@PCRonline Thank u genius!

Michael Megaly 的头像
Michael Megaly4 个月前

@PCRonline The genius people are rhe ones who invented it my friend